Abstract
Normalized intra-access pressure (PIA), expressed as the access pressure/systemic blood pressure, detects venous outlet stenosis and correlates with access blood flow. General use of (PIA) is limited by time, special equipment needs, and cost. We therefore correlated pressure measurements from the venous drip chamber (PDC of Fresenius H-machines and from an external transducer, P tau, for blood flows (BFR) of 0 to 400-500 ml/min. Measurements were conducted 2-3 weeks apart in a cohort of 33 patients. PDC = -21 + 1.28 P tau; PDC = P tau = 75 mmHg at BFR = 146 ml/min. The major determinant of P tau at BFR = 0 was access type and venous outflow problems. The difference between P tau and PDC (delta = offset) was 17 +/- 1 mmHg (range, 2-43); delta correlated with the height difference between the two sites. Differences in systemic blood pressure, zero calibration, and hydrostatic pressure accounted for 90% of the variance between replicate measurements of PDC. Detection of outlet stenosis was compared by using PIA calculated from P tau and from PDC + 17. Only three of 66 measurements using the latter produced misclassification, and never on replicate measurements. P tau and PDC measurements in 62 additional patients showed a persistent offset of 17 mmHg. The authors conclude that PDC at BFR = 0 can be used to monitor prospectively prosthetic bridge grafts for stenosis as long as the offset for a particular dialysis machine is determined.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.