Abstract

A single-step high performance liquid chromatography (HPLC) method for vitamin D quantitation in blood serum is described. Methanol/hexane extraction, silica Sep-Pak purification, differential dissolution, filtration, and the use of a precolumn in place of an injector sample loop allow for using only one (reversed phase) HPLC step instead of the usual two. Detection limit with the HPLC system used was about 1 μg/L of serum. Recovery of vitamin D 3 added to serum ranged from 90% to 109%. Intra and interassay coefficient of variation values were 8.1% and 15.4%, respectively.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.