Abstract

In this study, scaled scan band was developed to provide size-specific dose estimation (SSDE) values based on head circumference of patients undergoing computed tomography (CT) scans. The scan band was tested in 40 consecutive head CT examinations. The accuracy of the specialized scan band method was determined by comparing SSDEband with SSDE293,forehead, SSDEmean and SSDEcenter. SSDE293,forehead was used as the control value. The results of the linear fit of SSDEband, SSDEmean and SSDEcenter against SSDE293, forehead, were R2=0.958, R2=0.984 and R2=0.936, respectively. There was no significant difference between SSDEband, SSDEmean and SSDEcenter for SSDE293,forehead. Use of the proposed scan band method makes it possible to accurately determine the required radiation dose before a CT examination is performed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.