Abstract

Electron magnetic circular dichroism (EMCD), the electron wave analog of x-ray magnetic circular dichroism (XMCD), allows for the element specific measurement of the spin and orbital magnetic moments with up to nanometer resolution. However, due to dynamical diffraction effects, the signal-to-noise ratios of EMCD spectra are often very low. We describe a simple set of rules, how to set up a geometry for a classical EMCD experiment on an arbitrary crystal structure to get a maximum dichroic signal. The procedure is based on an evaluation of the structure factor and extinction distances. Proof-of-concept simulations and experiments on a FeGe crystal present a successful test of these guidelines.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call