Abstract

Cardiac output by the thermodilution technique was measured by a new No. 2 Fr. transthoracic (2F-TT) thermistor catheter placed at cardiac operation into the pulmonary artery directly through the right ventricular outflow tract. Cold (0 degree C) 5% dextrose in water (D5W) was used as indicator and injected through a percutaneously placed central venous pressure (CVP) catheter in the jugular vein. Comparison to the No. 7 Fr. Swan-Ganz (7F-SG) catheter demonstrated a close correlation (r = 0.87) and almost identical mean thermodilution cardiac output values during 530 determinations in 10 patients. No difficulty was experienced in insertion or removal of the 2F-TT catheter and no bleeding complications were noted. Experiments in six dogs showed that variation in position of the tip of the CVP catheter within the superior vena caval venous system and right atrium was not a critical factor in measurement of thermodilution cardiac output. The thermodilution cardiac output technique in general and the ease of insertion, as well as the small size of the 2F-TT catheter, should make this method especially advantageous in infants and small children.