Abstract

Purpose: This article’s objective is to develop a simple methodology deliver a uniform radiation dose to the wall of a narrow peripheral artery for preventing restenosis using a high-dose-rate (HDR) 192Ir remote afterloader. Methods and Materials: Based upon published two-dimensional data such as anisotropy factors of an HDR 192Ir source calculated from the Monte-Carlo method, arterial wall doses at a close range from an HDR source may be easily calculated using the special formula suggested in Task Group Report No. 43 published by the American Association of Physicists in Medicine. An optimization procedure was used to calculate the optimized dwell times for delivering a uniform dose along arterial walls for various arterial diameters and lengths of lesions. Results: Based on lengths of the stenosis and diameters of arteries or angioplasty balloons, a set of simple look-up tables for optimal dwell time intervals of endovascular radiation treatment have been developed for the MicroSelectron HDR remote afterloader. Conclusion: Doses for endovascular irradiation have been accurately calculated with anisotropy factors. For delivering uniform doses along the arterial wall, a set of look-up tables listed for optimal dwell times is available for the HDR remote afterloader.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.