Abstract
The construction of a simple disposable instrument for the measurement of cervical dilatation is described. The instrument consists of two clips, a measuring tape and a flexible tube. The measuring tape is fixed to one clip and threaded through the flexible tube, which is fixed to the other clip. The clips are attached to opposing cervical rims at 3 o'clock and 9 o'clock and the differential displacement of the marking on the measuring tape and free end of the flexible tube is measured. Laboratory measurements were performed in order to determine the error of measurement. The absolute error of measurement with this instrument is 1.5 mm. Clinical measurements were performed on 12 parturients with a single foetus in vertex position and 3 to 4 cm cervical dilatation. The results of measurements were within ±1 cm error margin of manual assessments. No adverse effects were noted due to the use of the instrument. The instrument did cause minor discomfort to patients during application.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.