Abstract
Arapid, simple and sensitive high-performance liquid chromatography method was developed for determination of ciprofloxacin in plasma by means of ultraviolet detection. Ofloxacin was used as an internal standard and separation carried on a Novapak C18 column using a mobile phase of 0.01 M phosphate buffer (pH =2.6): methanol (82:18 v/v). Extraction of drug was performed from plasma by liquid-liquid extraction and the average recovery was 78.2% The assay is precise, with inter-assay coefficient of variation of 6.70 % at 0.25-8 kg/ml (n=3). Using UVdetection at 277 nm the detection limit for ciprofloxacin was 20 ng/ml of plasma and the mean extraction recovery was 78.2 %. Short elution time, using UVdetector and usage of ofloxacin as internal standard are advantages of this method
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.