Abstract

Various means may be used to improve the detectability of an x-ray diffractometer, such as using pulse-height selection, etc. One method of increasing detectability is that of using a helium path instead of an air path for the x-rays. A graph comparing the detectability of diffraction-line intensities in a helium path to that in an air path for various wavelengths is given in Fig. 1. Chromium K alpha (CrKα) radiation is used for greater dispersion of large d spacings such as are encountered in many organic compounds.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.