Abstract
For the quantification of azasetron in rat plasma samples, a column-switching HPLC method was developed and validated. Following dilution of plasma samples with mobile phase A (17 mM potassium phosphate buffer (pH 3.0)) and simple protein precipitation by addition of perchloric acid (60%), the mixture was directly injected onto the pre-column. After endogenous plasma substances were eluted to waste, the analyte was transferred to the trap column by switching the system. Then, the analyte was back-flushed to the analytical column for separation with mobile phase B (a 22:78 v/v mixture of acetonitrile and 17 mM potassium phosphate buffer (pH 3.0)) and detected at 250 nm using a photodiode array detector. A linear standard curve was obtained in the concentration range of 10-800 ng/mL with the correlation coefficient (r) of 0.9998. The intra- and inter-day precision and accuracy values for azasetron were in the ranges of 0.3-12.9% and 89.7-101.4%, respectively. The method was valid in terms of specificity, precision, and accuracy. In addition, this efficient analytical method was successfully applied to determine plasma concentrations of azasetron following oral administration of azasetron at a dose of 4.0 mg/kg to rats.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.