Abstract
Zinc concentration in seminal plasma is one of the best biochemical parameters with which to assess prostatic function in the study of male infertility. Nevertheless, this parameter is not determined in many laboratories because of technical difficulties. In this paper a simple colorimetric method for seminal plasma zinc assay is presented. This assay is performed with only one reagent solution making deproteinization and extraction with organic solvents unnecessary. The method is based on the reaction between 1-(2-pyridylazo)-2-naphthol and zinc which yields a coloured complex, that is soluble in the reagent solution when dodecyl hydrogen sulfate sodium salt is present. The solubilized complex has an absorbance peak at 560 nm. The accuracy of this method is proved by statistical comparison with atomic absorption spectrophotometry. Furthermore, the method has a good precision and is free of interference.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.