A selective and highly sensitive isocratic high performance liquid chromatographic (HPLC) method is described for simultaneous determination of lactone and carboxylate species of topotecan, in plasma. The method utilizes a protein precipitation step with cold methanol (−20 °C) for sample preparation followed by separation on a Novapack C 18 column using ammonium acetate buffer, acetonitrile and triethylamine (84:16:1.5, v/v) containing tetrabutyl ammonium hydrogen sulfate (TBAHS) (2 mM) with a pH of 5 as the mobile phase. The eluted peaks were detected by a fluorescence detector was set at an excitation wavelength of 380 nm and an emission wavelength of 527 nm. The method was validated in the range of lactone and carboxylate forms of topotecan concentrations from 0.05 to 75 ng/ml. Intra- and inter-day precision expressed by the relative standard deviation was less than 8.50% and inaccuracy did not exceed 10% for lactone and carboxylate forms of topotecan. The limit of quantitation was 0.05 ng/ml using 0.50 ml plasma. Stability studies in plasma and plasma extract indicated that topotecan is stable for at least 2 weeks at −70 °C.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Similar Papers
More From: Journal of Chromatography B
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.