Abstract
Milnacipran, a serotonin and adrenaline reuptake inhibitor used to treat fibromyalgia and major depression, is difficult to quantify analytically at shorter UV wavelengths. This necessitates the development of a sensitive, precise and accurate colorimetric method for the determination of milnacipran hydrochloride in bulk and swellable matrix tablets. The drug was derivatized with ninhydrin reagent prepared by two methods using direct heating on flame and in glycerol bath at 110°C. The drug -(milnacipran)- ninhydrin complex (Ruhemman’s Purple) showed an absorbance maximum at 570 nm. The method exhibited linearity in a concentration range of 4 to 400 μg/mL with a correlation coefficient of 0.9999. As per ICH guidelines, the method showed good sensitivity (LOD = 0.550 μg/mL and LOQ = 1.667 μg/mL), intraday accuracy and precision (4.713%), and interday accuracy and precision (2.569%). The method was applied to determine the drug in swellable matrix tablets. The obtained results indicate that the proposed method of milnacipran hydrochloride determination in bulk and swellable matrix tablets is simple, accurate and precise, and can be used in less equipped laboratories.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.