Abstract

This paper presents a simple, rapid, and accurate voltammetric method for determination of boron in water and steel samples using a disposable pencil graphite electrode. The oxidation of Tiron in the boron-Tiron complex in phosphate buffer at pH 7.5 was measured as a response. Type and concentration of supporting electrolyte, pH, ionic strength, Tiron concentration, scan rate, step amplitude, pulse amplitude, and pencil grade were investigated as the parameters affecting the peak current. The limit of detection (3s) was estimated as 84 $\mu $g/L. The relative standard deviation of the method was 4.6 for 1 mg B/L (N = 7). The recovery results varied between 90% and 103% for water samples and between 94% and 108% for steel samples. The results were compared with those obtained from the ICP-OES method and no statistically significant difference between the methods was found.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.