Abstract

A high performance liquid chromatographic method for determination of moxifloxacin in human plasma was developed. The method involved deproteinisation of the sample with perchloric acid and analysis of the supernatant using a reversed-phase C 18 column (150 mm) and fluorescence detection at an excitation wavelength of 290 nm and an emission wavelength of 460 nm. The assay was specific for moxifloxacin and linear from 0.125 to 10.0 μg/ml. The relative standard deviation of intra- and inter-day assays was lower than 10%. The average recovery of moxifloxacin from plasma was 101%. Due to its simplicity, the assay can be used for pharmacokinetic studies of moxifloxacin.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.