Abstract

In an effort to develop a simple and mediator-free sensing media for urea, a polyaniline (PANI) and graphitized nanodiamond (GND) nanocomposite was prepared at room temperature via an in situ interfacial polymerization approach. The interaction of ions produced during the hydrolysis of urea with nanomaterial was utilized for the amperometric sensing of urea. The performance of this composite (GND/PANI) was evaluated for the detection of urea referenced to PANI. Accordingly, both platforms exhibited good linearity across a varying concentration of 0.1 to 0.9 mg mL-1 with limit of detection (LOD) values of 0.07 (PANI) and 0.05 mg mL-1 (GNDs-PANI). As such, both of them were able to detect urea well below the common range in real samples. Moreover, their sensitivities were also determined as 140.5 and 381.5 μA (mg mL-1)-1 cm-2 with a response time of 15 and 20 s, respectively. As such, the incorporation of GND with PANI was helpful toward enhancement of sensitivity.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.