Abstract

A simple procedure was developed for packing PicoFrit HPLC columns with chromatographic stationary phase using a reservoir fabricated from standard laboratory HPLC fittings. Packed columns were mounted onto a stainless steel ultra-low volume precolumn filter assembly containing a 0.5-microm pore size steel frit. This format provided a conduit for the application of the nanospray voltage and protected the column from obstruction by sample material. The system was characterised and operational performance assessed by analysis of a range of peptide standards (n = 9).