Abstract

Susceptibility artifacts and transmission radio frequency (RF) field (B(1) +) inhomogeneity are major limitations in high-field gradient echo MRI. Previously proposed numerical 2D spectral-spatial RF pulses have been shown to be promising for reducing the through-plane signal loss susceptibility artifact by incorporating a frequency-dependent through-plane phase correction. This method has recently been extended to 4D spectral-spatial RF pulse designs for reducing B(1) + inhomogeneity as well as the signal loss. In this manuscript, we present simple analytical pulse designs for constructing 2D and 4D spectral-spatial RF pulses as an alternative to the numerical approaches. The 2D pulse capable of exciting slices with reduced signal loss and is lipid suppressing. The 4D pulse simultaneously corrects signal loss as well as the B(1) + inhomogeneity from a body coil transmitter. The pulses are demonstrated with simulations and with gradient echo phantom and brain images at 3T using a standard RF body coil. The pulses were observed to work well for multiple slices and several volunteers.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.