Abstract

AbstractSilver microelectrode arrays are fabricated by photolithography for a one‐step analysis of H2O2 in low ionic strength samples. The effects of electrode length, width, band‐to‐band separation, connection height, and adhesion layer are evaluated. The developed sensor shows excellent repeatability (RSD=1.20 % (n=5)) and reproducibility (RSD=1.12 % (n=5)) with the linear range of 0.0–10.0 mM, the sensitivity of 9.84±0.34 μA mM−1, and the detection limit of 22.69 μM. The sensor has been successfully applied to detect H2O2 directly without the addition of supporting electrolyte in synthetic urine, tap water, drinking water, and milk samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.