Abstract

We retrospectively reviewed 42 patients who underwent resection of the distal ulna with implantation of a silicone rubber ulnar head prosthesis (45 wrists). Two prostheses were used: the original Swanson prosthesis, and a prosthesis of our own design. Follow-up X-rays showed migration or breakage of 63% of the prostheses. No statistically significant correlation existed between the quality of functional outcome and the integrity of the prostheses. There was no significant difference between pre-operative and post-operative range of motion for the entire group or between patients with broken or intact prostheses. Histological confirmation of silicone synovitis was documented in one patient who required implant removal. We suggest that destabilization and breakage of prostheses result from fatigue failure secondary to the torque generated at the distal radio-ulnar joint during repeated pronation and supination. Use of a silicone rubber ulnar head prosthesis following distal ulna resection is not recommended.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.