Abstract

Single- and double-lumen silicone rubber central venous catheters were subjected to in vitro destructive testing. Using this data we devised a clinical technique for the detection of central venous thrombosis incorporating mechanical injection of the catheters, serial filming, and magnification radiography. In 20 patients studied, thrombosis was detected in 12 and extravasation in 3. Two partially occluded catheters burst during forceful injection. Our technique demonstrated the presence and extent of thrombosis at the catheter tip more clearly than did other venographic methods and has been especially useful in assessing the results of thrombolytic therapy.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.