Abstract

In this paper, the authors report for the first time the physical cointegration of piezoelectric actuation and piezoresistive detection on resonating micromembranes dedicated to microgravimetric biosensing applications. The micromembranes are oscillated by a reverse piezoelectric phenomenon provided by a PbZrxTi1−xO3 46/54 thin layer. The oscillation amplitudes are read-out by measuring the resistance change of piezoresistors precisely located on the clamped edges of each micromembrane. The detection of the micromembranes' resonant frequencies is reported in air and deionized water. A dedicated electronic set-up operating the micromembranes in a closed-loop configuration is described. The set-up enables multiplexed tracking of four micromembranes' resonant frequencies in liquid media while enhancing the corresponding quality factors' values. Increases up to 11-fold of the micromembranes' quality factors in liquid is reported for the (0,1) vibration mode. A quality factor of up to 417 is reported in fluid.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.