An investigation was carried out into the chromatographic behavior, on a silica hydride-based phase and a comparator silica-based phase, of an important group of lipids endogenous to human plasma, which are associated with matrix effect and in the context of quantitative peptide analysis. The propensity for aqueous normal phase (ANP) retention on the silica hydride-based phase was strong and extensive in comparison with the silica-based comparator, and the lipophilic interferences in question were readily eluted using the ANP mode, a contrast to over-retention issues with accompanying implications for method ruggedness typically found with silica-based phases. The silica hydride-based phase, with ANP operation, offered selectivity conducive to rapid lipophilic interferent elimination and the bimodal retention involved in suitable gradient elution was appropriate for general peptide analytical application.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.