Abstract
A liquid chromatography electrospray ionization tandem mass spectrometry method with amoxicillin-d4 as the stable isotope-labeled internal standard for simultaneous quick detection of amoxicillin and clavulanic acid in human plasma was developed and validated. Chromatographic separations were performed on a Hedera ODS-2 column (2.1 × 150 mm, 5μm). The mobile phases for gradient elution were aqueous solution containing 0.2% acetic acid (AA) (mobile phase A) together with organic phase solution (acetonitrile and methanol mixed solution, mobile phase B). Mass spectrometry was performed using negative electrospray ionization in multiple reaction monitoring mode. The target fragment ion pairs of amoxicillin, clavulanic acid and amoxicillin-d4 were m/z 364.1 → 223.1, 198.1 → 135.9 and 368.1 → 227.1, respectively. The linear ranges of this method were 40-5,000 ng/ml for amoxicillin and 30-2,500 ng/ml for clavulanic acid, with coefficient of determination > 0.9900. This method validation included selectivity, standard curve, lower limit of quantitation, accuracy, precision, recovery, matrix effect (hemolytic matrix and hyperlipidemic matrix), carryover, stability, dilution reliability and incurred sample reanalysis study. A successful application of this method was realized in a pharmacokinetic study after administration of amoxicillin-clavulanic acid potassium granules.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.