Abstract

Impedimetric sensors can be used for the detection of biological targets, such as bacteria. In this paper, 3D finite-element simulations show that random distributions of bacteria on interdigitated microelectrode sensor surfaces restrict their limit of detection (LOD). We demonstrate the strong influence of the bacteria diameter, surface coverage and electrode geometry on both the sensitivity and the signal-to-noise ratio (SNR). Particularly, we optimize the microelectrode design to maximize the SNR and hence the LOD, and establish that the optimal electrode gap is approximately four times the bacteria diameter. Furthermore, we show that the electrode width and thickness should respectively be minimized and maximized in the considered range with regards to current technological limitations. The paper finally highlights and discusses, from the designer point of view, the critical design tradeoff between SNR and sensitivity maximization.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.