Abstract

Signal-to-noise ratios achievable with dispersive and nondispersive atomic fluorescence systems are compared. In this comparison particular attention has been devoted to considerations affecting optimization of the dispersive system with low and high background flames. Relative signal and noise values are reported for Hg and Fe atomic fluorescence with dispersive, solar-blind nondispersive, and filter nondispersive systems with C2H2-air, separated C2H2-air, H2-air, and H2-O2-Ar flames. It is concluded that, when due consideration is given to optimization of the dispersive system for atomic fluorescence measurements, use of a nondispersive system will not generally provide an improvement in the measurement signal-to-noise ratio and may result in a decreased signal-to-noise ratio with an atomizer of even moderate background emission, such as the separated C2H2-air flame.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.