Abstract

The efficiency and quality of three-dimensional Fourier transform (3DFT) magnetic resonance imaging were evaluated and compared with those of 2DFT imaging. For a fixed imaging time and number of sections, 3DFT imaging with conventional spin echoes always had a worse signal-to-noise ratio (S/N) than did 2DFT imaging, when both were optimized with respect to choice of repetition time (TR). With partial-flip magnetic resonance (MR) imaging, S/N was nearly equal for 2DFT and 3DFT imaging when both were optimized with respect to TR and flip angle. 3DFT imaging can have cross-section artifacts that exceed those of 2DFT imaging. For very thin sections these artifacts may be lessened, and 3DFT imaging can achieve this with smaller gradient pulses. Over-all, 3DFT imaging was found to be advantageous only for the very-thin-section imaging and in combination with partial-flip MR imaging.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.