Abstract

Most of the power in the signals from microcalorimeters occurs at relatively low frequencies. At these frequencies, typical amplifiers will have significant amounts of 1/f noise. Our laboratory systems can also suffer from pickup at several harmonics of the AC power line, and from microphonic pickup at frequencies that vary with the configuration of the apparatus. We have developed some optimal signal processing techniques in order to construct the best possible estimates of our pulse heights in the presence of these non-ideal effects. In addition to a discussion of our laboratory systems, we present our plans for providing this kind of signal processing in flight experiments.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.