Abstract
T1 and T2 weighted signals derived from various radiological contrast media were studied during MRT spin-echo sequences. In addition, the interaction between radiological contrast media and Gadolinium-DTPA concerning T1 signals was evaluated. Ionic (ioxitalaminic acid) and non-ionic radiological contrast media (Iopromid, Iotrolan) were used in diagnostic concentrations. Measurements were carried out with a superconductive magnet of 1.5 Tesla. Radiological contrast media produced significantly higher signals than a physiological sodium chloride solution in T1-weighted spin-echo sequences. Even small amounts (15% of total volume) of radiological contrast media during T1-weighted spin-echo sequences led to a significant reduction (about 25%) of the signal intensity of a 2 mM Gadolinium-DTPA solution. This may lead to diagnostic problems, as was shown in a series of 25 MR arthrograms of the shoulder. It is recommended than an interval of at least 6 hours should elapse between the use of a radiological contrast medium and an MRT examination.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: RoFo : Fortschritte auf dem Gebiete der Rontgenstrahlen und der Nuklearmedizin
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.