Abstract

ABSTRACTIntroductionThe sigma metric offers a quantitative framework for evaluating process performance in clinical laboratories. This study aimed to evaluate the analytical performance of automated analysers in haematology laboratories, using the sigma metric to choose the best analyser as an internal reference analyser.Materials and Methodsinternal quality control (IQC) data were collected for 6 months from SNCS, and the sigma value was calculated for 9 haematology analysers in the laboratory.ResultsFor the normal control level, a satisfactory mean sigma value ≥3 was observed for all of the studied parameters of all automated analysers. For the low control level, platelet (PLT) count by Instrument (Inst.) G performed poorly, with a mean sigma value <3. Inst. H, with all parameters’ sigma values >4, performed best and was chosen as the internal reference analyser.ConclusionThe sigma metric can be used as a guide to choose the QC strategy and plan QC frequency. It can facilitate the comparison of the same assay performed by multiple systems.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.