Abstract
The routine use of veno-venous bypass during the anhepatic phase of orthotopic liver transplantation is controversial. Decreased shunt flows (1.5-3.0 litre min-1), as reported in the literature, may explain the lack of beneficial effects on outcome. We have studied the influence of bypass flows on caval pressure gradient (CPG) and renal perfusion pressure (RPP) in 45 patients undergoing orthotopic liver transplantation using a portofemoro-subclavian veno-venous bypass system. Mean shunt flow was 3.63 litre min-1. Second-order polynomial regressions best described the relationship between shunt flow and CPG (r = 0.674), RPP (r = 0.727), and cardiac output (r = 0.602). Shunt flows less than 3.0 litre min-1 failed to normalize CPG and RPP, whereas flows greater than 5.0 litre min-1 did not substantially improve haemodynamic state.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.