Abstract

BackgroundWe tested the repeatability of myocardial blood flow (MBF) quantified using 82Rb with and without motion correction (MC) and with arterial input functions estimated from left ventricle (LV) and atrium (LA). MethodsTwenty-one patients referred for clinical 82Rb PET/CT underwent repeated rest scans in a single imaging session. Global MBF was quantified using three different assessments by two operators: (1) automatic processing without MC and LV arterial input function (AIF), (2) with MC and LV-AIF, and (3) with MC and LA-AIF. Inter-scan and inter-operator repeatability were tested using coefficient of variation (CV). ResultsMC with LV-AIF did not change MBF (no MC: 1.01 ± 0.30 mL/min/g vs MC with LV-AIF: 1.01 ± 0.29, P = 0.70), whereas MC with LA-AIF showed significantly lower MBF assessments (0.95 ± 0.28 mL/min/g, P = 0.0006). We report significant improvement for test-retest reproducibility for global MBF following MC (CV; No MC: 16.0, MC (LV-AIF): 9.2, MC (LA-AIF): 8.8). Good inter-operator repeatability was observed for LV-AIF (CV = 4.7) and LA-AIF (CV = 5.6) for global MBF assessments. ConclusionsMC significantly improved the test-retest repeatability between operators and between scans. MBF obtained after MC with LV-AIF were comparable, whereas MBFs after MC and LA-AIF were significantly reduced.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.