We describe an interferometric technique for the demodulation of serial fibre Bragg grating sensor arrays, yielding absolute measurement of the individual grating mean wavelengths. The composite beam reflected from the array illuminates a scanning Michelson interferometer but, in contrast to spectral measurement by Fourier Transform Spectroscopy, our technique requires an OPD scan far shorter than the coherence length of the grating reflections. The technique is based on the high-resolution measurement of the phases of the complex analytic signals and of the relationship of these phases to interferometric delay. The analytic signals are derived via the Hilbert transform, allowing frequency domain filtering of individual signals within the signal processing. The technique has yielded a resolution of 0.007 nm for an OPD scan of 1.2 mm.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.