The aim of this work is to design a set of gradient coils with an optimal geometry for magnetic resonance microscopy studies. Designs for a three-axis gradient coil system particularly suited for studies with small radiofrequency coils are presented. The novel geometry involves a planar section with concentric return paths to keep the coil short. Reduction of the external field has been attempted by varying the positions of the return paths using a simulated annealing algorithm. A biplanar version of x- and z-directed prototype gradients was built and tested. A 2D-MR image of a grid phantom has been obtained on a 7-T MR instrument to demonstrate the theory. A three-axis set used as a surface gradient set has also been built and used to obtain high-resolution MR images.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.