Abstract

The output video signal from three X-ray fluoroscopy machines was recorded with a storage oscilloscope to calculate the dynamic range of their image intensifier television camera chains. The two MERATE MTH90 and MGR50hf machines employed the same imaging chain but the response of the former to exposure was significantly lower because of a three channel light distribution optical system which was installed but not used. The CGR Stenoscope machine was found to have a very narrow dynamic range to the extent that it was not usable. Additional measurements on the CGR machine showed that the image intensifier brightness control mechanism would produce stabilized brightness in less than 0.4 s and it proved to be faster than the manual selection of exposure parameters. The quality of the image stored in the memory device on the same system was compared with the original by subtracting the two video signals on the oscilloscope and thus acquiring a quantitative measure of their difference.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.