Abstract
The objective of this cross-sectional study was to validate the accuracy of a hand-held electronic on-farm test (Precision Xtra) for quantifying the blood β-hydroxybutyrate (BHBA) concentration in dairy goats. A total of 114 dairy goats from 3 commercial herds were sampled once for blood in the jugular vein between 1mo before and 2mo after parturition. Blood samples were centrifuged to harvest serum and sera were sent to the Animal Health Laboratory of the Université de Montréal for quantification of BHBA concentration (gold standard). Laboratory BHBA values were between 0.1 and 3.7mmol/L. Precision Xtra values were compared with gold standard values; Pearson correlation coefficient was 0.98 and coefficient of determination was 0.95. Overall, these results suggested that Precision Xtra provides excellent accuracy for measuring blood BHBA concentration in dairy goats compared with the gold standard test.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.