Abstract

We provide the first demonstration of an acousto-optically modulated quantum cascade laser (AOM QCL) system as a diagnostic for combustion by measuring nitric oxide (NO), a highly regulated emission produced in gas turbines. The system provides time-resolved broadband spectral measurements of the present gas species via a single line of sight measurement, offering advantages over widely used narrowband absorption spectroscopy (e.g., the potential for simultaneous multispecies measurements using a single laser) and considerably faster (>15 kHz rates and potentially up to MHz) than sampling techniques, which employ fourier transform infrared (FTIR) or GC/MS. The developed AOM QCL system yields fast tunable output covering a spectral range of 1725–1930 cm−1 with a linewidth of 10–15 cm−1. For the demonstration experiment, the AOM QCL system has been used to obtain time-resolved spectral measurements of NO formation during the shock heating of mixture of a 10% nitrous oxide (N2O) in a balance of argon over a temperature range of 1245–2517 K and a pressure range of 3.6–5.8 atm. Results were in good agreement with chemical kinetic simulations. The system shows revolutionary promise for making simultaneous time-resolved measurements of multiple species concentrations and temperature with a single line of sight measurement.