Abstract

A shipboard fluorometric flow analyzer has been developed for near-real-time, high-resolution underway measurement of ammonium in seawater. The fluorometric method is based on the reaction of ammonium with o-phthaldialdehyde (OPA) and sulfite. The reagents used in this method have been modified to suit seawater analysis. This method shows no refractive index and salinity effect from seawater samples. The potential interferences in seawater have been studied, and their effects have been reduced. The instrument response is linear over a wide range of ammonium concentration. The limit of detection of 1.1 nM was estimated in laboratory using ammonium standards prepared in distilled water. It should be noted that application of this method to low-level ammonium measurement requires a correction of interference species, such as amino acids. The sample throughput is 3600 h(-1). The system can be used for both freshwater and seawater samples and has been used to monitor the distribution of ammonium in Florida coastal waters around an oceanic wastewater outfall.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.