Abstract

Instruments suitable for removing calculus, plaque and necrotic cementum, which hinder normal periodontal reattachment, are extremely important for successful therapy. The test was carried out in order to see if a standard scaler used for ultrasonic tartar removal maintains its physical features when sharpened. 6 scalers, compatible with piezo-electric generators, were tested and 2 diameters, A and B, weights and resonance frequencies were measured. Scalers no. 2, 3, 4, 5, were sharpened and was used as control. All measurements were taken again at the end of the test. Statistical analysis reveals significant variations, after sharpening, of A diameter (t = 4.14 greater than 3.55, p less than 0.01), B diameter (t = 5.34 greater than 3.355, p less than 0.01) and resonance frequency (t = 2.82 greater than 2.306, p less than 0.05); in contrast, there is no significant change of weights (t = 1.17 less than 2.306, p greater than 0.05). These results suggest that sharpening slightly modifies the physical features of ultrasonic tips, so that sharpening ultrasonic scalers, from a physical point of view, can be carried out, paying attention not to damage the water cooling system.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.