Abstract

Early graft occlusion after coronary artery bypass grafting may have deleterious consequences. We routinely use transit-time flowmetry after termination of cardiopulmonary bypass, and we report five cases of early graft failure detected by the flowmeter. Electrocardiographic (ECG) changes were seen in only one of these five cases, and none of the patients had low cardiac output or other signs of graft failure at the end of the operation. The cause of graft failure was tagging in one case, rotation of internal mammary artery grafts in two and kinking of vein grafts in two cases. All errors were corrected, and control flowmetry showed normal flow rates after correction. Flowmetry takes less than 10 min, even with multiple bypass grafts. Based on our results, we advocate routine quality control with flowmetry after termination of cardiopulmonary bypass, since ECG changes are insufficient as checks of flow in bypass grafts.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.