Abstract
Endothelial decompensation is a serious complication of phakic intraocular lens (pIOL) implantation and is a major concern during the postoperative period. We report 3 eyes in which the same foldable angle-supported pIOL was implanted to correct high myopia. Rapid and severe postoperative endothelial cell loss occurred in all 3 eyes. An over-sized pIOL that induced excessive vaulting into the anterior chamber was the main risk factor. In 2 eyes, the pIOL was explanted uneventfully; 1 eye required Descemet's stripping automated endothelial keratoplasty because of total endothelial decompensation. These cases illustrate the importance of accurate sizing of foldable angle-supported anterior chamber pIOLs to avoid excessive vaulting. They also highlight the importance of regular follow-up and preventive pIOL explantation as soon as significant endothelial cell loss is detected.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Similar Papers
More From: Journal of Cataract and Refractive Surgery
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.