Abstract
Our study describes a 55-year-old man with a total hip prosthesis (ceramic femoral head and polyethylene [PE] inlay). After a fall, the ceramic head broke into several pieces, and a subsequent revision surgery with metal femoral head and PE inlay was performed. Three months later, the so far healthy patient complained of multiorganic symptoms. The subsequent clinical examinations revealed hypothyroidism, peripheral neuropathy, and cardiomyopathy for the first time. In a second revision surgery, a massive deterioration of the metal femoral head by overlooked particles of the broken ceramic head was found. The cobalt concentration in blood exceeded a never before observed level. After replacement of the scraped femoral head, the cobalt concentration decreased considerably, and most of the symptoms disappeared.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.