Abstract

Methods for correcting errors in solar absorption FTIR spectroscopy using the Sun as the light source were developed. A novel camera-based solar tracker is presented, providing robust tracking accuracies better than 30 arc sec. A method for correcting mis-sampling artifacts in the interferograms is shown. A portable and robust low-resolution FTIR spectrometer was developed for measuring total column CO2 values. An agreement of 0.1% was reached compared to a co-located TCCON reference instrument.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.