Abstract

<p class="Abstract">Measurements of mechanical quantities are often carried out with transducers with a bridge output. The output signals are conditioned using bridge amplifiers. If dynamically changing quantities are going to be measured traceably, the bridge amplifier must be calibrated dynamically.</p>This paper describes a dynamic bridge amplifier calibration setup based on the new PTB dynamic bridge standard. The calibration is carried out by the synchronous sampling of the bridge amplifier output voltage and a reference signal provided by the calibrated dynamic bridge standard. The dynamic bridge standard enables calibrations in a frequency range from DC (static calibration) up to 10 kHz. An overview of the different measurement uncertainty contributions is given, and the first measurement results show good agreement with a previously established measurement setup.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.