Abstract
A triple-to-double coincidence ratio (TDCR) counter was recently constructed at IRA-METAS for liquid scintillation based primary activity standardisations. A description of its optical chamber, efficiency change tools, photomultiplier tubes (PMTs) and electronics is given. This TDCR system was validated by measuring several standard solutions of beta emitters including 45Ca, 14C, 63Ni and 3H. The activity concentrations, obtained from these measurements and efficiencies computed with a FORTRAN code we developed for symmetric and asymmetric PMTs, agree with the certified values within uncertainties.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.