Abstract

A case is presented in which an elevated serum titanium level was used to make the diagnosis of a failed metal-backed patellar component. The preoperative serum titanium level was 536.8 ppb, which was 98 times higher than the patient's previous level (taken 1 year earlier, when he was asymptomatic) and 2 orders of magnitude higher than the expected level with a well-functioning implant of this type. Revision surgery confirmed that the polyethylene portion of the patellar component had worn through, leaving the titanium portion of the patellar implant to articulate with the femoral component. Wear-through was not evident on preoperative radiographs or clinical examination. As knowledge about the expected ranges for serum metal ion levels after total joint arthroplasty continues to increase, the diagnostic utility of serum metal ion testing in the evaluation of joint arthroplasty function will continue to improve.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.