Abstract

The goal of this study is to evaluate serum levels of chromium (Cr), cobalt (Co), and titanium (Ti) within the first two years following total hip arthroplasty using a Ti modular neck system. Twenty-five patients were randomized to a metal-on-metal (MoM) bearing with an all CoCr shell, and the remaining 25 received a metal-on-polyethylene (MoP) with a Ti shell. Serum levels demonstrated increases for Cr, Co, and Ti at 1 year (P < .001). MoM had similar Ti levels to MoP hips at 1 year (P=0.11) but lower at 2 years (P=0.03). Results suggest that the passive corrosion (i.e., chemical, pitting, and crevice corrosion) of exposed non-articular metal surfaces may be a greater source of ions than the neck-stem or head-neck interfaces.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.