Abstract
A new pair of Epstein-Barr virus ELISAs (Biotest Anti-EBV VCA IgG and VCA IgM ELISA) was evaluated for usefulness for routine diagnosis of acute EBV infections. The ELISAs are based on two viral capsid antigens (VCA), p23 (BLRF2, full-length) and p18 (BFRF3, carboxy-half), that are combined by autologous gene fusion. In total, 179 sera were tested in direct comparison with classical VCA immunofluorescence assays (IFA). With the help of clinical data and additional reference serology, i.e., heterophile antibodies, anti-EA IgG (IFA) and anti-EBNA-1 IgG (ELISA), the patients were divided into the following categories: seronegatives (46), acute primary infections (67), previous infections (39), suspected reactivations (20) and constellations with intermediate serological patterns (7). The VCA IgG and VCA IgM ELISAs showed overall agreement to IFA of 95.0% and 94.4%, respectively. The calculated analytical performance (sensitivity; specificity) of VCA IgG and VCA IgM was 94.0%; 97.8% and 97.1%; 96.5%, respectively. A certain delay in seroconversion of anti-p23-p18 IgG may account for a significant difference in sensitivity of the VCA IgG ELISA between primary (88.4%) and previous infections (100%). In summary, the new recombinant VCA ELISAs yielded good correlation to VCA IFA and in combination with EBNA-1 IgG allow rapid, sensitive, and specific diagnosis of infectious mononucleosis or EBV immune status in general.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.