Abstract
AbstractA sequential voltammetric procedure for the determination of uranium, cadmium and lead was investigated at an ex situ bismuth film electrode (BiFE). First, the adsorptive stripping voltammetry was applied to assay the U(VI)‐cupferron complex in the differential pulse mode (detection limit of 1.0 µg L−1, 200 s accumulation time). Through the manipulation of the same aliquot of the sample, efforts were made to quantify cadmium and lead by square wave anodic stripping voltammetry. Detection limits of 2.03 µg L−1 for Cd (II) and 2.43 µg L−1 for Pb (II) were calculated (100 s accumulation time). The methodology was successfully applied to phosphate fertilizer samples after open vessel wet decomposition (HNO3/H2O2). The following value ranges were evaluated: U (VI) 37.2–150 mg kg−1, Pb (II) 78.3–204 mg kg−1 and Cd (II) 44.1–71.6 mg kg−1. Validation was performed by using the standard reference materials SRM‐695 – phosphate fertilizer – and SRM‐1643e – water.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.