Abstract

An easily automatable sequential flow-injection-pervaporation method is proposed for the photometric determination of methanol and iron in vinegar. The method is based on separation of the methanol from the sample matrix by pervaporation followed by its oxidation to formaldehyde with permanganate, decolouration of the latter with S(2)O(5)(2-), and subsequent reaction of formaldehyde with p-rosaniline to yield a violet reaction product with maximum absorption at 567 nm. Iron is determined by an existing method based on reaction with thiocyanate in acidic medium and monitoring at 508 nm. After optimisation by either the univariate or multivariate approaches, as required, the linear range was established for methanol (4-1000 mg L(-1)) and iron (0.18-20 mg L(-1)). The proposed method was then compared with reference methods for methanol and iron in terms of repeatability (2.452 mg L(-1) and 0.245 mg L(-1), respectively), reproducibility (4.435 mg L(-1) and 0.356 mg L(-1), respectively), detection limit (LOD=82 and 0.234 mg L(-1), respectively), and traceability. The throughput was nine samples per hour.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.